SN Equipment Code Name of Equipment Specification

viraginityfumblingSoftware and s/w Development

Nov 2, 2013 (3 years and 8 months ago)

1,344 views

SN

Equipment
Code

Name of
Equipment

Specification

1.


MCHB001

Cardioc scope
with central
monitor

Cardio scope with monitor

Patient Multi
-
Para Monitor

Specifications


Should have facility to display 12 leads ECG, RR, HR, SpO2, NIBP, 2 Temperature, as standard

parameters, with built
-
in rechargeable battery back up of 1 Hrs.




Display : Should have minimum 14” or above color TFT touch screen display.


Should be modular monitor with modular interchangeable facility .



Should display at least 6 waveforms of sel
ected parameters simultaneously.



Should have Nellcor/maismo/ FAST technology to sense the SpO2 in hypotensive, shivering &
motion condition.



Should have Oscillometric Technology for measurement of NIBP with Auto, STAT and manual
modes.




Should have d
ifferent patient type selection (From neonate to adult).



The pulse rate should be displayed either with ECG or SpO2.



Should be able to analyze arrhythmia & ST segments for all the leads.



Should have facility for displaying multi
-
screen configurations
.




Should be able to store & display 72 hours or more of graphical trends of all parameters.



Should be suitable for monitoring adult, pediatric & neonatal patients.



Should be able to give visual & audible alarms with three levels of volume adjustment

on violation
of any monitored parameter.



Should have connectivity to Central station.



Quoted model should be FDA/ CE / TUV approved. Authentic Certificate has to be submitted.

• optional : should have moduler upgradability for Etco2 ,multigas facility

and cardiac output , BIS ,
EEG , thermal recorder ,



Should be supplied with:

• 5 lead ECG cable


1

• SpO2 (Adult) probe with extension cable


1


SpO2 (Pediatric) probe with extension cable


1


• NIBP Cuff with extension cable for Adult, Pediatric
and neonate


1 each

• Temperature (Skin) probe


2



disposable IBP transducer 2 no.

• Earthing Cable


1


• Power Cord


1


• Operating Manual
-

1


2.


MCHB002

AED Plus
Defibrillator

Should be operable by mains and battery

Mains
-
100
-
240volts AC (Auto ra
nging )Frequency 50/60 hz

Battery capable of usage for atleast 90 minutes or 40 discharge of 200 joules, Battery should be
rechargeable

Monitor/Display should have high resolution colour TFT screen size minimum of 6.4inch

Defibrillator should be Bi pahsic
truncated exponential optimised terminbating current impedance
compensated

Energy (External ) 1
-
10,15,20,230,50,70,100,120,150,170,200 joules (Internal) 1
-
10,15,20,30,50
joules

Should have AED Mode and Manual Mode

Should have facility for syncronised & man
ual defibrillation/cardioversion

Should be low energy bi phasic defibrillator and monitor with recorder having capability to revert
all arrhythmias within a maximum energy of 200 joules

Should monitor ECG through paddles, pads and monitoring electrode and
defibrillate through
pads and paddles

Should print the ECG on thermal paper of min 50 mm width

Shock delivery via multifunction defibrillator electrode pads or paddles

Should have external paddles with paddle contact indicator
-
for good paddles contact bot
h adult
& Peadiatric should be available recorder should have events summmary facility for recrding &
printing atleast 250 events & 50 wave forms Should capable of printing reports on events
summary,configurations , self test battery capacity etc.

should h
ave facility for self test/check before uses & setop functions record mode should be
manual& automatic non invasive pacing should have noninvasive extranal transtutanious pacing
by extranal paddles for both adult & pediatric mode of pacing both demand & s
ynchronous pulse
oximeter should have spo2 facility with adult spo2 finger probe,peaditric spo2 probe & spo2 ear
probe

NIBP monitering should have nibp monitoring facility for both adult & peadiatric


it should manual & automatic mode

there should be provi
sion for upgradeable to ETCO2

there should be optional USB port for data transfer & storages

system configuration , accessories , spares & consumables (A) defibilator (B) paddles Adult &
Pediatric patient cable, power cord, rechargeable battery , spo2 fing
er probe adult & pediatric,
spo2 ear probe 2nos each, nibp hose / tubing, nibp cuff adult & pediatric 2nos each, external
pacing pads adult & pediatric 2pair each.

should be US FDA or CE approved product

3.


MCHB003

Pulse
Oximeter

System should have:

1

High
-

resolution color TFT display of minimum 6” (diagonally) screen size.

2

Should display SpO2 values, PR value, and plethysmograph along with perfusion level indicator.

3

Should display online short trend for at least an hour.

4

SpO2 Range 1
-
100 %

5

Alarm Limits:


S
pO2

50
-
95 % (Low), 55
-
100 (High)

6

Adult to neonatal applications.

7

24 hours data storage facility in Tabular & Graphical mode with zoom facility.

8

Must have Graded & Color coded alarms (2 stage)

9

Alarm recall facility to view minimum last 15 alarms should be p
ossible

10

Inbuilt thermal array recorder facility.

11

Can be used as transport and ambulatory monitor.

12

External monitor or RS 232 option should be provided.

13

Mains and removable Lithium battery operation.

14

Conforming to International standards


IEC 60101


1


2
, CE.

Should be supplied with:
Reusable SpO2 finger sensor

4.


MCHB004

Laryngoscope
with blades

The set of four satin blades, includes:



Curved macintosh type, size: 1, 2 & 3.



Straight miller type, size: 1.



Spare bulbs
-

4 pcs.



Stainless steel satin handle us
es two alkaline batteries (medium size, 1.5v battery).

5.


MCHB005

Suction
Machine
heavy duty
Electric

Specifications :

Portable suction unit: size 38x17x28.5cm, weight 5.1kg

Description (e.g.: model Vario A/C, manufacturer Medela)

Technical specificati
ons: SUCTION MACHINE with pump:

-

Power supply: 230
-
240V/50Hz

-

Vacuum capacity: 18 litres/mim

-

Maximum depression:
-
75kPa (
-
563mmHg)

Vacuum is created by a plastic piston and cylinder system, with four vacuum
-
creating modules
(Quatro
Flex™ technology).

The membrane vacuum regulator permits accurate vacuum settings.

Double overflow
-
protection system (bottle and pump).

Working temperature range: +5 to +40°C.

SAFETY DEVICE:

The safety device protects the pump against overflow: suct
ion is interrupted when foam or liquids
reach the safety level in the small jar. All components are reusable and can be steam sterilized at
134°C.

A collector system with a single
-
use protection device has been added on request of the ICRC

COLLECTOR SYSTE
M:

Bottle or jar: 1 litre, graduated PSU (polysulphone)

Lid: Includes 1 overflow device, a 6x10mm diameter conical connector and 2 clips

Connection tube: Silicone tubing, diameter 12x7mm, length 25cm, and 2 connectors: 1 straight
and 1 curved 90°.


Suction tube: Reusable silicone tube: diameter 7x12mm, length 200mm.

Single
-
use tube: diameter 7x12mm, length 180cm.

Filters: Bacterial filters, single
-
use.

Unit presentation: Delivered with the following components:

-

1 suction machine and pump


-

2 one
-
litre bottles or jars, with lids

-

1 safety device

-

4 overflow spares

-

2 lid clip spares

-

2 suction tubes, non
-
sterile

-

2 suction tubes, sterile

-

1 spare fuse

-

1 user and maintenance manual, multilingual

Packaging: Box: 400x400x47
0mm

The following must appear on the pump packaging:

-

Manufacturer’s name and address.


-

Item designation, serial number, CE mark and reference number of notifying body.

Outer packaging: Manufacturer’s name; product name and serial number.

Other:

Complies with Medical Device Directive 93/42/EEC, class risk IIa or equivalent.

Manufacturer’s quality management system complies with ISO 13485.

Accessories
-

Spare parts

(Manufacturer Medela: model Vario 18 CA, 600.2601)

Code + description Pictures
Comments

XSEQSUCC3B


(suction, elec., vario) BOTTLE,

1L, psu, autoclavable

Liquid collector

-

volume content 1 litre

-

graduation scale 100ml

-

material polysulphone

sterelisable at 134°C .

XSEQSUCC3BL

(s
uction, elec., vario) LID w/ conn.6
-

10mm, + overflow device


Components:

-

1conical connector Ø 10
-
14mm

-

1 overflow protection device

-

2 blue plastic clip, for lid fixationXSEQSUCCM30B
LC

(suction, elec., b30/vario) CLIP for lid

clip, plastic, blue, for lid fixation

XSEQSUCCM30BLO

(suction, elec., b30/vario)

OVERFLOW protection device

Bottle /jar overflow protection device,

white

XSEQSUCC3S


(suction, elec., vario) SAFETY

DEVICE, for pump, Ref 2001733


Components

1 Botte/ jar 250ml, polisulphone

1 Lid with patient connector Ø 6
-
10mm, two


clips and one overflow protection device

2 Connecting tube, silicone



size: external Ø 12mm, internal Ø 7mm,


length 30cm + 2 connectors


All components are reusable and

autoclavable at 134°c

XSEQSUCCM30T

(suction, elec., b30/vario) TUBE

SUCT., d.12x7mm/2m, silicone.

SuctionTubin
g (between operation field and

suction machine) reusable

Size: external Ø 12mm, internal Ø 7mm,

length 2m

Material silicone,

Autoclavable at 134°C

MCTDTSUCC25

TUBE, SUCTION MACHINE, CH

25, 1.8m, sterile, s.u.

Suct
ionTubing (between operation field and

suction machine) single use

Size: external Ø 12mm, internal Ø 7mm,

length 1.80m, with manual vacuum control

connector

XSEQSUCC3U

(suction, elec., vario) FUSE T

800 mA, dia.5x20mm




Fuse T 0.8 A Ø 5.0 x 20 mm

Accessories not included in standard version 2009, can be ordered separately as option ICRC
Code + description Pictures Comments

XSEQSUCC3A

(suction, elec., vario) BAG, transport

-

Made of strong resistant


waterproof textile

-

Multi pocket bag allowing to

store accessories

XSEQSUCC3R

(suction, elec., vario) TROLLEY, w/

2 fix.rails, Ref 0770511

Material: stainless steel

2 standard rails 10x25mm

4 swivelling antistatic castors, 2

with breaks

XSEQSUCC3RC

(suction, elec., vario, trolley) CLAMP

HOLDER, Ref 0770821


Support fixation clamp, for rails

10x25mm

XSEQSUCC
3F

(suction, elec., vario) FILTER, antibacterial/overflow, s.u

Overflow protection /bacterial

filter is compulsory for the

suction model not equipped with

reusable safety device (for

ICRC all XSEQSU
CC3

delivered before 2009)

Note:

When ordering technical spare parts aleays mention, suction apparatus model, and serial
Number

6.


MCHB006

Elect. Cautery
Machine with
Bipolar facility

Salient Features



Powerful and relatively charr free bipolar coagulatio
n



Step less variation of output power



Audio and visual indication for foot switch activation



Highly portable unit



Autoclavable accessories.



Inbuilt voltage stabiliser



Special power transformer for non
-
interference with other monitoring equipment

Optional




Various shapes of Bipolar forceps suitable for specialized surgical needs

StandardAccessories



Colour coded single pedal footswitch with receptacle



Insulated Bipolar forceps with connection cable (Autoclavable)

Technical Specifications



Bipolar Coagulation P
ower : 100 Watts



Rated frequency : 1000 KHz



Power supply : 230 Volts, 50 Hz



Line Voltage Fluctuation Compatibility



Working range : Line Voltage from 150 to 270 Volts

PhysicalDimensions



Unit size : 304 x 180 x 123 mm



Unit gross weight : 4.5 Kgs

7.


MCHB007

BA
TTERY
OPERATED
DRILL SYSTEM
WITH
VARIOUS
ATTACHMENT


Cannulated AO Drill


Fully cannulated (3.2mm) with variable


Speed (0
-
1200 rpm)


Separate trigger drive for clock wise &


Anti clock wise direction

Double loose pin fully autoclavable

Radiolucent drive &should accept all

Attachment

1.

Jacobs chuck with key


cannulated with variable opening from 0
-
6.5mm

2.

Quick coupling
-

For drill, tap, screw, medullary and Acetabular reamer attachment.

3.

Quick coupling

-
For ‘K’wire upto3.2mm.

4.

Quick coupling
-
For DHS/DCS reamer.

5.

Radiolucent drive for drilling of distal interlocking holes.

6.

Drill bit for radiolucent drive of 2 to 4 mm diameter.

7.

Oscillating saw attachment with variable amplitude 2.5

o

to 5
o
with oscillating frequency from
12500/min to14000/min

8.

Oscillating saw blade with cutting thickness from .5 to 1.5thickness.

9.

Reciprocating saw attachment with blade of 1.1 to 1.2mm thickness.

10.

NiMH/NiCd battery with rating at least 1.5 mAh. Separate batt
ery for drill and saw with one
spare battery each.

11.

Battery charger with capacity to charge 4 batteries simultaneously and input voltage 220
-
240v.

12.

Flexible cannulated reamers
-

various sizes (5 mm to 14 mm, in 1mm increment)

Oscillating drill attachment f
or converting continuous rotatory movement to back and forth


8.


MCHB008

Image
Intensifier
with C
-

Arm &
Memory &
Specifications

Fully microprocessor based system with feather touch panel,image intensifier, microprocessor
based control wit
h touch panel

Printing
facility

System description :
-
fully microprocessor based system with feather touch panel

Image Intensifier
:Thomson / Toshiba 6 imaging system with High performance capability.

Generator
:


X
-
Ray

Tube
-
Head
:

Collimated X
-
Ray output through im
ported x
-
ray tube with following features:

Dual focii having 0.6mm and 1.8 mm Focus

Anode heat storage capacity 30KJ, Anode cooling rate 250w

Power Input:
220
V
, 1 Phase, 16 Amps, 50 Hz.

Generator Rating
:

Microprocessor based control with feather Touch Panel
.

Generator Surgipix IIC unit delivers Excellent radiography.

Radiographic output from 40mA to 100mA

Fluoroscopy 0
-
5 mA (as per kvp selection) variable in steps of 0.1 mA.kvp variation from


minimum


40 kvp


to 100 kvp in steps of 1 kv

Line voltage compen
sation:

Stabilized 230V output at supply line voltage of 180V to 270V.

Additional short circuit protection with auto trip Feature.

Control features:

Fully Microprocessor based digital voltmeter for indicating Mains supply


voltage.

Digital kvp meter displa
y.

Digital mAs display.

Digital mA meter display.

Digital radio timer display.

Foot
-
switch for spot operation

Feather touch keyboard with appropriate led indications.

III.
Mechan

Smooth manual
-
control of front and back wheel steering for easy operation.

Mot
orised C
-
arm control with smooth up and down movement.

Well balanced C

Smooth manual horizontal travel, panning & C Rotation.

Trolly for T.V. Monitor with easy movement.

C
-
ARM
Features
:

Rotation :


360
0

*


Source to I.I. Distance



:


90 cms.

*


Arc orbital movement


:


110
0

*


Panning


:


25
0

*


Horizontal travel


:


23 cms.

*


Vertical motorised


:


35 cms.

*



Movement Detachable

*


Cassette holder


:


10 X 12


or 8 X 10 film.

TV
Monitor
: 15/17 B/W, 625 lines, 50hz, 230v Ac with high resolution display capability 2 numbers
monitor supplied.

* IMAGE MEMORY:

*


32 Frame Memor
y With Dual Video Output And Frame Averaging.

*


Step
-
Wise Image Integration.(2/4/8/16)

*


Image Reversal Positive/Negative

*


Pulsed Fluroscopy

*


Image Resolution (512 X 512) Pixles.

*


Dynamic Storage Of Imag
e





Digital Display For Image Selection

*Image Memory:




Memory a nd Da t a ba s e

1.


Acquisition is up to 25 frames/sec (selectable in increments of 1).

2.


50 cine
-
loops (1 cine
-
loop = 200 Images) are stor
ed in temporary memory before saving
it in permanent storage. Hence an addition of about 10,000 images to the total memory mentioned
in next point.

3.


Total storage: 25,000 Images + 405 cine
-
loops or 105,000 Images for permanent storage
in

basic system.

4.


Storage of cine
-
loops with frame
-
by
-
frame viewing and images.

5.


Auto run of the cine
-
loop.

6.


Patient information database to store, manage, retrieve Patient, Procedure, Image and
Cine
-
l
oop information stored in the system.

7.


Image Mosaic or Multi
-
map to allow thumbnail viewing of images for quick access.

Query the database using key words

8.


Post Image Processing like images rotation, zooming, panning,
measurements,
annotation, edge sharpening, pseudo colouring etc.


9.


MCHB009

Ventilator

High End Ventilator Specifications:

1.

Description of Function

1.1


ICU ventilators should provide artificial respiratory support to the critical patients in all the
types of
Intensive Care Units with altitude compensation for volume and BTPS correction for monitoring.

2

Operational Requirements

2.1


It should be microprocessor Controlled ventilator with integrated facility for Ventilation monitoring
suitable for Neon
atal (including premature) to adult ventilation.

2.2


The unit should be external compressor based for precise gas delivery ( not a turbine / piston/
blower based) .It should have proximal flow sensor for neonatal patient category only .
Demonstration of
the equipment is a must.

3

Technical Specifications

3.1


Hinged arm holder for holding the circuit

3.2

Should have Colored Touch screen, 12 Inch or more

3.3


It should have facility to measure and display ( inbuilt )

a) Mainstream End tidal CO2 with Capn
ography ( Optional) with monitoring of PeSO2, Vmin VCO2,
Curve Co2/ time, Alveolar ventilation , expire Co2 ( VTCo2) ,Vhigh and low alarms etc

b) 3 waves
-

Pressure and Time, Volume and Time and Flow and Time.

c) 3 loops
-

P
-
V, F
-
V, P
-
F with facilit
y of saving of 4 Loops for reference.

d) Graphic display to have automatic scaling facility for waves

e) Status indicator for Ventilator mode, Battery life, patient data, alarm settings, clock etc

3.4


should have Trending facility for 72 hours

3.5


should

have Automatic compliance & Leakage compensation for circuit and ET tube

3.6


should have Following settings for all age groups.

a) Tidal Volume 5 ml to 2500 ml

b) Pressure (insp) 2
-

100 cmH2O

c) Pressure Ramp/ Flow patterns

d) Respiratory Rate 1 to 150 b
pm, Insp. Time 0.1 to 3 sec, I : E Ratio 5:1 to 1:599

e) Insp. Flow ( resultant) 0.2 to 180 LPM, continuous Flow 2
-
40 lpm ( in TCPL ode)

f) CPAP/PEEP 0
-
50 cmH2O

g) Pressure support 2
-
100 cmh2O

h) FIO2 21 to 100%

i) Pause Time 0 to 2 sec

j) Flow Trigger 0.
2 to 15 lpm . Pressure Trigger 0.5 to 20 cmH2O

k) Expiratory trigger 5
-
80
-
% of flow

3.7


should have ,monitoring of the following parameters

a) Airway Pressure (Peak & Mean)

b) Tidal volume (Inspired & Expired)

c) Minute volume ( Expired)

d) Spontaneous Mi
nute Volume

e) Total Frequency

f) FIO2 dynamic

g) Intrinsic PEEP and PEEPi Volume (or trapped Volume)

h) Plateau Pressure

i) Resistance ( Rinsp & Rexp)& Compliance ( Cdyn & Cstat)

j) Use selector Alarms for all measured & monitored parameters

should have
following modes of ventilation

a) Volume controlled

b) Pressure Controlled

c) Pressure Support

d) SIMV (Pressure Control and volume control) with pressure support

e) CPAP/PEEP, PSV + assured tidal volume

g) Advanced mode like pressure controlled volume
guaranteed


( PRVC)

h) Non Invasive ventilation

i) MMV+PSV

j) APRV

k)Special Neonatal modes
-

TCPL, SIMV +TCPL + PSV, N
-
CPAP

( with continuous flow) , PRVC,APRV,and above mentioned mode

3.9


should have Apnea /backup ventilation

3.10


Expiratory block sho
uld be autoclavable and no routine calibration required

3.11

should have ,monitoring of the following parameters

a. Intrinsic Peep & Intrinsic PEEP Volume ( Trapped Volume)

b. Occl
usion Pressure( P0.1) , Max Inspiratory pressure (Pi max)

c. Non
-
forced Slow Vital Capacity , physiologic Dead space, RSBI, Imposed work of Breathing (
WOBi), Expiratory Time constant


( Tcexp)

d. Facility to calculate lower and upper inflection point ( P/
V Flex points)

e. Facility for ET Tube and Patient circuit compensation

3.12


should have Nebuliser with capability to deliver fine particle size of to be used in On line

3.13


should have Ideal Body Weight facility

3.14


should integrated Battery back u
p for minimum 2 hour for main unit

3.17


RS 323C interface for communications with networked devices.


4

System Configuration Accessories, spares and consumables


4.1


ICU Ventilator with trolley
-

01

4.2


Adult , Pediatric , Neonatal autoclavable silico
ne patient breathing circuits
-
02 each

4.3


Reusable and autoclavable Flow sensor and exhalation valve
-

2 each

Proximal Flow sensor for neonatal use
-

5 nos

Hinged Support Arm
-

1no

Air and Oxygen Hose
-

each 1 no


Optional Accessories

4.4


Medical Air

Compressor

4.5


Reusable Masks (Small, Medium, Large) with each machine.
-

02 sets ea

4.6


Humidifier
-
Servo controlled with digital monitoring of inspired gas temperature complete with
heating wire
-
01 ( Optional)

4.7

EtCo2 Sensor ( reusable)
-
1 no

Inlet r
equirement

5.1


Power input to be 220
-
240VAC, 50Hz

5.2

Gas input( air and oxygen)
-

50
-
100 psi

6

Standards, Safety and Training

6.1


Should be US FDA / CE approved product

6.2

Demonstration of quoted equipment model is a must.

6.3

Should have local se
rvice facility .The service provider should have the necessary
equipments recommended by the manufacturer to carry out preventive maintenance test as per
guidelines provided in the service/maintenance manual.

6.4

Comprehensive warranty for 1 year and provi
sion of AMC for next 5 years.


10.


MCHB010

Multipara
Monitor



Should have facility to display ECG, RR, HR, SpO
2,

NIBP, single Temperature, as standard
parameters, with built
-

in rechargeable battery back up of 1 hrs.



Optional upgrades: dual channel Invasive

blood pressure, EtCO2 measurement, multi gas
monitoring and recorder with built
-

in rechargeable battery back up of 1 hrs.



Display: Should have minimum 15” or above color TFT display.



Should display at least 10 waveforms of selected parameters simultaneou
sly.



Should have Nellcor technology to sense the SPO2 in hypertensive, shivering & motion
condition.



Should have Oscillometric Technology for measurement of NIBP with Auto, STAT and manual
modes.



Should have different patient type selection (from neonate t
o Adult).



The pulse rate should be displayed either with ECG or SpO2.



The respiration rate should be calculated through Impedance method



Should have single skin temperature monitoring either in
0
Celecius or
0
Fahrenheit.



Should be upgradeable to dual skin t
emperature monitoring.



Should be able to analyze arrhythmia (18 types) & ST segments for all the leads.



Should have facility for displaying multi
-

screen configurations (Standard, NIBP review, BIG
numerics, Short trend, 7 leads, 12 leads, OXY
-

CRG, Other
bed).



Should be able to store & display at least 96 hours of graphical trends of all parameters.



Should be suitable for monitoring adult, pediatric & neonatal patients.



Should be able to give visual & audible alarms with three levels of volume adjustment o
n
violation of any monitored parameter.



Should have connectivity to Central station through Wireless Station.



Nurse call function should be included as standard configuration.



Facility of bed to bed remote monitoring when connected via Ethernet to central
monitoring.



Should have software’s like Drug dose calculation and OXY


CRG as standard.



Quoted model should be FDA,CE, ISO/ TUV approved compulsory. Authentic Certificate from
has to be submitted.



Should be supplied with:

o

5 lead ECG cable


1

o

SpO2 (Adult)

probe with extension cable


1

o

NIBP Cuff with extension cable for Adult, Pediatric and neonate


1 each

o

Temperature (Skin) probe


1

o

Earthing Cable


1

o

Power Cord


1

o

Operating Manual


1

11.


MCHB011

Syringe pump

Syringe Pump

Programming systems

a)

Flow rate

b)

F
low rate and volume

c)

Flow rate and volume

d)

Volume over time; the pump automatically calculates the Flow rate

Automatic syringe
-
size detection

Adjustable occlusion pressure and ABS

Simple and safe BOLUS function

Full rechargeable battery

Full information abou
t the infusion status

12.


MCHB012

Infusion
pump

Infusion Pump

Flow Rate range:

a)

0.1 to 99.9 ml/h (in 0.1 ml/h increments)

b)

100ml/h to 999ml/h (in 1ml/h increments)

Accuracy: 5% with assigned IV set

Total Volume Range: 1 to 9999 ml or no limit

Purge Flow Rate
: 700ml/h (or selectable)

Alarms and Indicators:

a)

Air in Line

b)

Infusion complete

c)

Occlusion (Up/down)

d)

Empty container

e)

Low battery

f)

Door open

g)

Tubing misleading

h)

Other device malfunctions

Function:

a)

Keep
-
sensor

b)

Drop sensor

c)

Tubing clamp mechanism

d)

Broad range of de
livery rate settings

e)

Volume memorize function

f)

Selectable one shot purge

g)

Easy replacement of battery & convenient battery indicator

h)

Free drop protection finger

i)

User selectable setup (IV set, sound, occlusion, etc.)

Power Requirement: AC 100 to 240 V, 50/60h
z, External 12V DC, Power consumption: Less than 30
VA

Classification: Class I, Type BF, IPX 1

Battery Source: Rechargeable Nickel
-
Metal Hybride

Battery can run the pump continuously for about 3 hours at an operating rate of 25ml/h

Accessories: AC power,

IV Sets, Operating Guide
132)
)Infusion pump


13.


MCHB013

Oxygen
concentrator

Input Voltage
-

120 VAC ,

Input frequency
-

60Hz ,

Average power consumption


350 Watts

Oxygen concentration (at 5 LPM)


93% ,


Litre flow
-

31lbs (14kg) ,

Oxygen percentage
indicator alarm level


low ,


Oxygen
-

82% ,Very low oxygen


70% ,


Operating Humidity
-

95% ,

Concentrator Accessories


,

Compressor intake filter , Micro disk filter ,

Humidifier Connector tube

2 ,



Flow meter
-

2, for two patients at the same
time133)

14.


MCHB014

Electronic
Pneumatic
Tourniquet
(fully
automatic
micro
controller
based, with
dual cuff for
orthopaedic
operations

-
The electronic tourniquet should have digital display of time duration and cuff pressure.

-
It should have feather touch

electronic switches.

-
It should have battery backup so that pressure is maintained even in case of power failure.

-
Online decrease or increase of pressure should be possible.

-
Should have positive locking connector.

-
Should have noiseless operation.

-
Sh
ould afford quick pressurization of cuffs.

-
Should not require external high pressure.

-
Should have Dual cuff operation .

-
There should be an autoclavable silicon cuff and coiled expandable puncture resistant P U tubing.

-
An extra cuff and tubings should b
e provided.

-
The cuff pressure range shoul be from 10 to 450 mm of Hg.

-
Pressure regulation:+/
-
10 mm of Hg.

-
Timer could be set to wide range of 1 hour to 9 hours.

-
There should be an audible alarm on timer reaching set value.

-
It should have a memory fun
ction i.e.pressure set in last session is stored and displayed when
machine is switched on again.

-
Power supply:230 V,AC.,50+/
-
10%.,Built in battery for 3 hour back up,should work on generator
without the need of a voltage stabilizer.

-
Should have 1 year w
arrenty againt manufacturing defects including cuffs.

-
Should be light weight as little as 4 kilos.

-
SIZE: Length215mm,Height 100mm,Width 190mm.

-
Cuff sizes:Large: 33”,5”


Big : 24”,4”


Medium: 18”
,3”.


Small : 11”,3”


Pediatric: 8” 2”.

-
Should able to bear temperatures as high as 110*c for autoclaving.

15.


MCHB015

Blood gas
analyzer with
expenses of
reagents

Technical Specifications:



Equipment interface :
240 X 128 dots LCD display with soft key pad



Analysis principle : Direct "flow Thru" electrodes



Sample volume : ≤200µ



Sample type : Whole blood, serum, plasma.



Sample application : Sample cup, capillary or collection tube



Throughput : 65 samples / hour



Calibration : Automatic two
-
point calibration with additional option for carrying out one
-
points an
d two
-
point calibration manually



Pre adjusted standby mode with automatic calibration steps



Quality control : 2 different levels of quality controls



Printer : Built in stationary head graphical thermal printer



Storage Capacity : 1000 samples



Interface port

: RS 232



Power : 110 230V AC, 50/60 Hz



Dimensions : 452(H) X 452 (W) X 500(D) mm approx



Equipment weight : 15 kg approx



Operating environment : 15º to 30 ºC , 20% ~ 80% RH, 86 ~ 106kPa

16.


MCHB016

Oxygen
Cylinder kit

One Aluminium Cylinder with Regulator, P
ressure Gauge, Flow Meter and humidifier with
mask Oxygen in




Lts.
-

100 Lts (0.1 m
3
)



Weight of kit
-

1.3kg.



Flow Rate
-

0
-
4 Ltrs/Min

Three Aluminium Cylinder with Regulator, Pressure Gauge, Flow Meter and humidifier with
mask Oxigen in



Lts.
-

300 Lts(0.3
m
3
)



Weight of kit
-

3.7kg.



Flow Rate
-

0
-
4.5 Ltrs/Min (Each cylinder flow is set on 1.5L/Min)

Aluminium Oxygen cylinder with 5 Cylinder with Regulator, Pressure Gauge, Glow meter and
humidifier and mask. Stand with tubing



Oxygen
-

500 Ltrs (0.5 m
3
)



Weight
-

5.7 kg



(Flow is set on 1.5L/Min)

Features:



Easy to Operate



Quick Application



Portable Light Weight



Refillable



Economical and Long Lasting



Quality Product



Can be a life saver



An excellent life
-
long healthy gift



Service to the patient

17.


MCHB017

Operation
t
able Hydrolic
with
radiolucent
top with all
ortho, and
neuro,
attachments

Salient features SI
-
112 Surgical Operation Table Hydraulic Side end Control:



Simple action hydraulic pump controlled by a foot pedal



Fully S.S. covered base mounted on heavy duty cas
tors with auto floor locking system the
detachable leg section provides easy lithotomic positioning



Five sections S.S. top with large perineal cutout for drainage tray, built in kidney bridge



The improved hydraulic system provides smooth movement and the s
peed of lowering the
table is permanently set



The table is finished with extra head, permanent finish oven baked paint and all the accessories
are bright chromium plated


Standard accessories:



Anesthetic frame



Padded side supports



Intravenous arm board



Pad
ded shoulder supports



Lithotomic leg holder with straps, footrest with S.S. top


Optional accessories:



Water proof rubber mattress



Radio
-
graphic top



Neurosurgical attachment



Geopel type lithotomic clutches with straps



Orthopedic attachment


Technical dat
a:



Length of the table top: 190 to 200 cm



Maximum height: 95 + 5 cm



Longitudinal tilt: 30 +1



Width of the table top: 50 cm



Hydraulic lift: 22 cm



Minimum height: 80 + 5cm



Lateral tilt: 20+2

18.


MCHB018

Emergency
crash cart

1. Overall size shall be more than 9
00mm L x 500mm W x 1500mm H.

2. The crash cart should be made of 25.4mmx18G Stainless steel tubular frame work.

3. Shall have Epoxy / Anti
-
Microbial powder paint inside and out

4. Should have dual push handles on either side

5. Should have S.S. shel
ves, six colored removable bins & two polystyrene lockable storage units
with three drawers each.

6. Facility to carry ECG Monitors, Defibrillators etc on open areas at top centre and bottom

shelves.

7. Should have Stainless steel saline rod fixed wit
h.

8. Two accessory mounting brackets to mount accessories anywhere without the need of
prethreaded holes.

9. Crash cart should be mounted on 12.5 cms dia non
-
rusting swiveling castor wheels. Two

having locking arrangement.

10. Oxygen cylinder stand

epoxy powder coated, on one side



19.


MCHB019

Emergency
Hydrolic
trolley



Overall approx Size: 1800mm x 675mm x 750mm



Frame work made of rectangular M.S. tubes



Stainless Steel top in 3 sections



Trendlenburg positions adjustable by Screw/pneumatic gas spring

system.



Backrest section adjustable on rachet



Leg end section can slide under the main section



Height adjustable a pair of knee crutches



Pre Treated and Epoxy powder Coated.

20.


MCHB020

Auto analyser

Description of Function :

Used for the extraction and det
ermination of hydrogen cyanide (HCN) in

the total particulate matter (TPM) and gaseous phase of tobacco smoke by

an automated continuous flow analyzer. This method is applicable for the

testing of cigarettes and bidis .Also for determination of Nitrate

,

Reducing Substances, Reducing Carbohydrates in Tobacco

Operational Requirement;

Autoanalyser complete with colorimeter, auto sampler and other accessories

suitable for HCN estimation.

Technical Specifications:

Should comprise of the following:


Autosampler.

• Peristaltic Pump

• Nitrate Manifold designed for AutoAnalyzer used.

• Single Channel Colorimeter equipped with 15 mm flow cell and 550

nm filter /RP18 5 µm, 250 mm X 4 mm Chromatographic HPLC

Column

• Sampler
-

Random access sampler wi
th throughput of more than 150

samples in cups or tubes.

• Manifold
-

Should include applications for Hydrogen Cyanide

measurements.

• Flowcell
-

Should be having bubble through the flow cell operation.

• Colorimeter should be suitable for Hydrogen Cyan
ide analysis

Environmental factors:

1. The unit shall be capable of being stored continuously in ambient temperature

in the range of 0
-
50 deg C and relative humidity in the range of 15
-
90%.

2. The unit shall be capable of operating continuously in ambie
nt temperature in

the range of 10
-
42 deg C and relative humidity in the range of 15
-
90% .

Power Supply:

1.Power input to be 220
-
240VAC, 50Hz fitted with BIS plug (ISI mark)

2.UPS of suitable rating with voltage regulation , spike protection and 60 minute
s power back up.

Standards, Safety and Training:

1. Should be FDA/CE/UL/BIS approved product.

2. Manufacturer should have ISO certification for quality standards.

3. On site Comprehensive training for lab staff and support services customer

satisfacti
on with the system.

4. Comprehensive warranty for 2 years and 5 years AMC after warranty.

5.Shall meet IEC
-
60601
-
1
-
2:2001(Or Equivalent BIS) General Requirements of

Safety for Electromagnetic Compatibility or should comply with 89/366/EEC; EMCdirective

6. The bidder shall assure supply of spares for a minimum period of 10 years from the date of
supply

Documentation:

1. User/Technical/Maintenance manuals to be supplied in English.

2. Certificate of calibration and inspection.

3. List of Equipments ava
ilable for providing calibration and routine Preventive

Maintenance Support as per manufacturer service/ maintenance manual.

4. List of important spare parts and accessories with their part number and

costing.

5. Log book with instructions for daily, w
eekly, monthly and quarterly maintenance checklist. The
job description of operators and service engineers should be clearly spelt out.

6. Compliance Report to be submitted in a tabulated and point wise manner clearly mentioning
the page/para number of
original catalogue/data sheet. Any point, if not substantiated with
authenticated catalogue/manual, without which it will not be considered.

7. Current user’s / performance list to be provided

21.


MCHB021

X
-

ray
machine
portable
100MA

Product Description

1
00mA medical mobile x ray machine YSX0410


GENERAL


1. Used in ward and emergency treatment room for radiography.


2. Combined X
-
ray generator.


3. Single focus, bridge rectification.


4. It is precise safe, reliable and flexible.


5. Remote control device
(control range≥ 5m)is used.

X
-
ray generator

Stationary anode, single focus and bridge silicon rectification.

The x
-
ray generator is an oil
-
immersed self
-
cooled unit.

Max rated capacity

Current

15
mA

30
mA

60mA

100mA

Voltage

90k
vp

90k
vp

90kvp

80kvp

Tim
e

6.3s

6.3s

4.0s

3.2s

Power supply

Voltage

180
-
240V

Frequency

50Hz

Capacity

Not over than 7KVA

Intemal resistance

Not larger than 1Ω

Voltage adjustment range
of x
-
ray tube

50
-
90KV

Range of the time

0.08
-
6.3s

Distance between the
focus to the grou
nd

700
-
1780mm

X
-
ray tube focus
transverse rotation

±60°

Specification of x
-
ray tuble

Model XD42.9/100 fixed anode, single focus 1.8/4.3mm

Transport
dimension(L×W×H)(mm)

1230×900×2050

Weight(kg)

Net:159/gross:256


22.


MCHB022

Ultra Sound
System 4 D
with
colour
duplex

Specifications

Imaging modes

l


B,2B,4B,B/M, M

l


CFM

l


PW Mode

l


Power Doppler/Directional PD

l


Duplex

l


Trapezoidal

l


Real
-
time 4D (Option)

l


Chroma B/PW

Probes

l


3.5MHz Convex probe

l


7.5MHz Linear probe

l


6.0 MHz Transvaginal pro
be

l


4.0MHz 4D Volume probe

l


7.0MHz Transvaginal probe(180°)

l


3.0MHz Micro
-
Convex probe (adult Cardiac)

l


5.0MHz Pediatric probe

l


Wideband, Multi
-
frequency

Image processing technology

l


THI on convex probe

l


Speckle Reduction Algorithm (SRA)

l


C
ompound technology

l


I
-
image: image optimization software

Measurement & Report packages

l


OB&GYN

l


Vascular

l


Urology

l


Small parts

l


Basic cardiac

Professional Clinical Applications

l


Abdominal

l


Obstetric

l


Gynecology

l


Vascular and Small parts

l


Pediatric

l


Urology

l


Musculoskeletal

Standard Configuration

Main unit, 15’’ LCD, 2 probe connectors, Hard disk, 2USB ports, Video out, TV out, LAN port, VGA
out


Options

l


3.5MHz Convex probe

l


7.5MHz Linear probe

l


6.0 MHz Transvaginal probe

l


7.0MHz Transvaginal probe(180°)

l


3.0MHz Micro
-
Convex probe (adult Cardiac)

l


5.0MHz Pediatric probe

l


4D package: including 4.0MHz 4D volume probe, 4D software and 4D hardware module

l


Video Printer: SONY UP897MD

l


PC Printer (any type)

l


DICOM 3.0

l


I
-
image: image optimization software

l


Biopsy kit: for convex, linear, TV prove

l


Foot
-
switch

23.


MCPTR023

PROSTHETIC
AND
ORTHOTIC
ELECTRIC
OVEN



Spare double walled, inside chamber made of stainless steel and out side made of mild steel



Heating is unifo
rm all around through high speed circulating fans



Toughened glass window on both doors



Supply with trays and inside stand for drape frames



Power supply 220
-
230volts single phase



60 x 75 x 50 cm


WXHXD



Temp 250
-
300*C digital Timers



Digital timer with alarm

24.


MCPTR024

ROUTER WITH
STAND &
ATTACHMENT



2.5 KW OR 2hp, 2800RPM with spindle, pedestal and starter.



Shaft length
-

350mm at inclination




Working height
-

850
-
900mm

25.


MCPTR025

HEAT PAN
WATER
HEATER FOR
LOW
TEMPERATUR
E
THERMOPLAS
TICS



65 x 10 x 50 cm(WxHxD) w
ith uniform heating



Body made of stainless steel both internally & externally



Digital indicator cum controller



Temp. range upto 80c
-
100c



Digital indicator cum controller

26.


MCPTR026

COMBINATIO
N GRINDER &
POLISHER



Grinding Stone: 20” Dia fitted on left sid
e



Shaft/spindle:2200
-
2400mm, long tapered fitted on right side



2Hp motor, rpm
-
2800
-
3000, Single Phase,

27.


MCPTR027

HAND
DRILLING
MACHINE



Drilling Dia
-

In 1
-
13mm



Pwer
-
50
-
600W



Speed
-
2500
-
3000rpm



Weight upto 2
-
2.5kg

28.


MCPTR028

HOT AIR GUN



Temp 100
-
600*C



Air fl
ow rate


350/550cpm



Power input
-

2000W



Weight 700gm
-
1kg

29.


MCPTR029

JIG SAW



Power input
-
500
-
600W



No load stroke
-

500
-
3000/min



Cutting depth in wood
-

80
-
90mm



Weight
-

1.5
-
2.5kgs

30.


MCPTR030

PILLAR
DRILLING
MACHINE



1” drill capacity



ISI mark drill chuck, arb
our and V belt



Adjustable drilling platform



4 drilling speed



single phase



RPM range 1500
-
2500

31.


MCPTR031

OSCILLATING
SAW



Ergonomic design with complete kit(Includes different attachment blades, spanner, fuse)



Circular blade with depth control gauge



Approxi
mate 25000
-
28000rpm



Effective for synthetic plaster and POP

32.


MCPTR032

ORTHOTIC
STAND FOR
MOULDING



Mild steel frame, 1000
-
1200 mm in height with platform for floor mounting



Suction mandrel made of mild steel heavy duty solid rod with 50 mm outer and 28 mm
inner
dia with brass ball valve.



Five planes adjustable alignment jig

33.


MCPTR033

HIGH
VACUUM
SUCTION
UNIT FOR
THERMOPLAS
TIC
MODULDING



Capacity
-
600
-
700mm hg, flutter free vacuum control knob



Pump type


double rotary vane pump



Jars are wide mounted 2 x 2
ltrs



Single phase 230
-
240 volts



Noise level very low

34.


MCPTR034

LINISHER
(BELT
SANDER)



Heavy duty steel body



Roller
-
150mm



Having duct, hose at lower hood to extract dust



Upper roller bearing with vertical and tilting adjustment for sander belt alignment



Be
lt size 80”x4”

35.


MCPTR035

ANVIL



Tip to tip length
-
15
-
20”



Height
-
5
-
7”



Face width
-
4
-
5”



Base width
-
5
-
7’



25 KGS, Mild steel

36.


MCPTR036

SELF
LEVELING
LASER WITH
STAND



Working range: up to 30
-
35ft



Leveling accuracy: +/
-
2/4
-
inches



Line accuracy: +/
-
3/32
-
in end to

end at 20’ perpendicular distance



Laser diode: two 635
-
670mm



Laser class: class 2M



Batteries: 3 “AA” alkaline batteries



Battery life: upto 50
-
60 hours of intermittent use

37.


MCPTR037

Bed side pedo
cycle



Should provide aerobic benefit of cycling to a bed ri
dden patient or those who cannot sit on a
static cycle.



Adjustable leg length & resistance level.



Self powered, smooth operation.



Straps to secure feet in pedals.



Digital/ LCD display of basic parameter showing speed, distance, time, calories, no. o
f cycle etc.

38.


MCPTR038

Kinesio sports
tape



Hypo allergic cotton fabric, latex free.



High adhesive quality (heat activated)



Should be ideal for sports soft tissue injury taping.



Water resistant should not get harden with change in temperature.



Each roll

-

width 2” length 5 meters.



Elasticity not less than 140%.

39.


MCPTR039

Therapy putty
(clay)



Should be ideal for hand function training.



Non oily, non toxic



Resistance category super soft to extra firm with colour code (at least 4
-
5 levels of resistance)




Should not fragment, separate, stain or stick to the skin



4
-
5 levels of resistance pack, weight 500gms each pack (2
-
2.5 kg)

40.


MCPTR040

Quadriceps
board



Made up of polished teakwood base with slots.



Two movable board with a hinge to allow quadriceps muscl
e exercise at various knee flexion
angle.



Base should be stable.



Smooth Edges.



Should be of ideal measurement for an adult patient.

41.


MCPTR041

Hand grip
exercise


Chrome plate frame rectangular in shape



Chrome plate springs

frame



Spring not less than 5 i
n numbers. Should be detachable to offer various level of resistance for
grip strengthening.



Wooden base for stability.

42.


MCPTR042

Power web
Rubber
webbing grip
exerciser



For finger hand grip & thumb exercise.



At least 3 level of different resistance.



1
4” ring diameter maximum.

43.


MCPTR043

Stability
trainer foam



Oval shape special stability trainer foam.



Should offer at least two or three levels of firmness.



Should provide progressive balance training for adult as well as pediatric age group

44.


MCPTR044

T
rampoline



Heavy duty frame and springs.



Minimum 40” diameter, 9” height



Vinyl reinforced stitching and commercial grade trampoline fabric



Should be ideal for pediatric use



Provision of hand rail for added stability



Max body weight allowed not less than
50kgs.

45.


MCPTR045

Exercise mat



Light weight high density foam with durable nylon fabric covering and reinforced stitching.



Reinforced handle at side for easy shifting



Peel & tear resistant, impermeable for fluid, crack resistant



Should allow easy cleanin
g



Mat size 120cm x 190cm x 10 cm

46.


MCPTR046

Exercise roll
(bolster)



Made up of high density foams with a solid plastic core covered with nylon fabric.



Peel & tear resistant



Impermeable for fluid



Crack resistant



Should allow easy cleaning



Stand on one e
nd & other end should must have provision for handle for easy shifting



Rack must for compact storage



Should be ideal for gross motor function, pediatric therapy & mat activity like rolling rocking.



Measurement As Follows



DIAMETER

LENGTH

15cm


6
0cm

20

90

25

90

30

90


47.


MCPTR047

Exercise
wedges




Made up of high density foams covered with nylon fabric.



Peel & tear resistant



Impermeable for fluid



Crack resistant



Should allow easy cleaning



Rack must for compact storage



Should be ideal for gr
oss motor function, pediatric therapy & mat activity.



Measurement as follows



WEDGE ELEVATION

BASE

15CM

60X70CM

20CM

60X70CM

25CM

60X70CM

30CM

60X70CM


48.


MCPTR048

Venile coated
iron
dumbbells



Should must be provided in pairs



Vinyl coated cast iro
n dumbbells



Thick smooth comfortable grip



Storage rack or trolley must.



Weight category (each in pairs) 500 gms, 1kg, 1.5kgs, 2kgs, 2.5kgs, 3kgs, 3.5kg, 4kg, 5kg

49.


MCPTR049

Soft ball
weight



Soft vinyl ball shaped weights



Should must be provided in pairs




Colour coded for different weights



Storage rack or trolley must.



Weight category (each in pairs) 500gms, 1kg, 1.5kgs, 2kgs, 2.5kgs, 3kgs.

50.


MCPTR050

Physio gym
ball

1.

Inflatable exercise balls

2.

Anti burst quality

3.

Non slip ribbed vinyl surface for extra

grip

4.

Max weight tolerance 200 kgs or more

5.

Storage rack or trolley must

6.

Sizes 45,55,65,75,85,95 cm

51.


MCPTR051

Peanut
shaped physio
gym ball.

1.

Should move in one direction only

2.

Should be more stable than conventional rounded gym ball

3.

Anti burst quality

4.

Non slip, ribbed vinyl surface for extra grip

5.

Max weight tolerance 300kgs or more

6.

Should be provided with a storage rack

7.

Sizes 85,100,120 cm

52.


MCPTR052

Continuous
passive
movement
lower limb

1.

Microprocessor controlled electronic circuit board.

2.

Knee flexio
n & extension 0°
-

130°

3.

Hip flexion & extension 0°
-

90°

4.

Ankle range 20°
-


-

40°.

5.

Flexion extension angle selection


one degree / selectable

6.

Treatment time


1
-

60 min

7.

Max patient weight allowed not less than 150 kgs

8.

Power source 24V adaptor

9.

Speed

should be adjustable

10.

Smooth operation

11.

Wired remote control for patient as standard accessory

12.

Adjustable calf & thigh length range

13.

Digital / LCD display showing treatment time, joint angle flexion & extension, speed.

14.

Emergency stop switch

15.

Light weight &

sleek design for easy transportation

16.

Cushioned surface with limb securing straps.

53.


MCPTR053

Rollator
walker junior
size

1.

Light weight walker

2.

Steel tubing frame

3.

Front wheels and back rubber tip ferrules.

4.

Adjustable handle bar length

5.

Height adjustment

24 to 30’’

54.


MCPTR054

Static bicycle
junior size

1.

Should be suitable for 5 to 15 year of age cycling

2.

Stable base

3.

Adjustable height, leg length

4.

Should offer various level of resistance

5.

Basic parameters like speed, distance, calories to be displayed in LC
D display.

6.

Straps to secure feet with the pedals.

55.


MCPTR055

Paediatric
walking
1.

Parallel bars with height (21
-
30”) and width (13
-
20”) adjustment

2.

Length 3 meters

parellal bar

3.

Wooden base platform for stable mounting

4.

Incline ends

5.

Hand rails of steel tubi
ng chrome plated

56.


MCPTR056

Treatment
table

1.

Made up of hard wood in natural finish with soft padding on top with Rexene cover

2.

Measurement 6ft x 3 ft x 3 ft

3.

One storage drawer underneath the top.

4.

Mounted on sturdily built hardwood

57.


MCPTR057

Motorized
treadmill

1.

Should be Ideal for both medical as well as sports fitness purpose.

2.

Adjustable ergonomic front and side bars.

3.

Easy to operate control panel display should be easily visible with LCD.

4.

Low mounting height

5.

Emergency stop button provision

6.

Programma
ble, in built TMT protocols

7.

Power full AC motor


2.1 Hp to 4 Hp

8.

Supply voltage


230 V +
-
10%, 50Hz

9.

Driving force


Frequency control rotator field electromotor

10.

Motor capacity


1500w to 1800w

11.

Max power consumption 10A

12.

Min
-
max speed 0.5 to 20 km/h

13.

Bel
t area


not less than 50x150cm

14.

Max client weight

not less than 130 kgs

15.

Inclination 0
-
25%

58.


MCPTR058

Heart rate
monitor

1.

Automatic calculation of training zone intensity with available and visual alarm

2.

Calorie expenditure with % of fat burn calorie

3.

99 la
ps /spilt stop watch with 1/100 sec resolution.

4.

Flexible display unit

5.

Provision for chest straps and wrist watch type sensor

59.


MCPTR059

Aesthesiomet
er
Monofilament
testing
apparatus

1.

Cutaneous sensation evaluation by log force calculation

2.

Should must cons
ist of glass fiber nylon monofilaments each precisely calibrated and of equal
length (not less than 10 in numbers from 1.65mm to 6.65mm thickness)

3.

Log force calculation and complete instruction manual should must be available

4.

Small & portable in size with

a protective case.

60.


MCPTR060

Pin wheel

1.

Stainless steel pin wheel for evaluation of cutaneous sensation along dermatome

2.

Length not less than 6”

3.

Smooth running pin wheel

4.

Pins should be of equal length with blunt end

61.


MCPTR061

Two point
discriminator

1.

S
tainless steel

2.

Compact design

3.

Straight, blunt tips

4.

Should allow precise measurement

62.


MCPTR062

Tuning fork

1.

For evaluation of vibration sense

2.

Minimum 30 cycle fork for low end testing

3.

Maximum 256 cycle for high end testing

63.


MCPTR063

Skin fold
callip
er

1.

For measurement of subcutaneous tissue fat

2.

Spring loaded arm for reproducible pinch

3.

Reading upto 60mm

4.

Light weight sturdy design

5.

Should must be provided with user manual

64.


MCPTR064

Stop watch

1.

Analog type

2.

Clear visibility

3.

Should provide information

about seconds, minutes, hrs.

65.


MCPTR065

Hydraulic
hand held
dynamometer
kit

1.

For evaluation of routine grip strength

2.

Analog type numerical display

3.

Should allow display of isometric grip strength from 0
-

25kgs

4.

Peak hold indicator needle

5.

Adjustment for gr
ip position

6.

Provision for pinch gauge 0
-
20 kg

66.


MCPTR066

Pelvic
inclinometer

1.

Should be ideal for pelvic tilt measurement

2.

Should measure distance + angle between two caliper ends

3.

Marks in cm or inches

4.

Chrome Plated Metallic Or Fiber Body Design

67.


MCPTR067

Inclinometer


1. Fiber body


2. Should be ideal for spinal range of motion measurement


3. Reliable and reproducible results







68.


MCPTR068

Artificial
anatomical
model of
following
joints

1. Should be ideal for demonstration/ teaching/patie
nt education purpose

2. Made up of fiber glass molded on base of wooden.

3. Should be maneuverable.

4. Required articulated joint models are as follows

1.

Right side articulated hip

2.

Right side articulated shoulder

3.

Right side articulated knee

4.

Right side artic
ulated elbow

5.

Right side articulated ankle

6.

Right side articulated wrist

7.

Right side articulated foot

8.

Right side articulated hand.

9.

Articulated whole spine

10.

Articulated cervical spine

11.

Articulated lumbo
-
sacral spine

12.

Articulated pelvis.

69.


MCPTR069

Ultra sound
the
rapy unit

1.

Dual head frequency treatment for 1 MHz.

2.

Mode : Continuous & Pulsed

3.

Output


15 Watts/ Power intensity


up to 3 Watts/cm2, Power source 220V AC / 50Hz

4.

Digital Display

5.

Timer


0
-
30 minutes with auto off & buzzer

70.


MCPTR070

Interferential
thera
py unit
1.

Output: 2 separate output channels I & II

2.

Output :0
-
90 MA

(ift)

3.

Output frequency :

channel I 4000 Hz


4250Hz depending on the base & spectrum control

channel II 4000 (constant)

4.

Selection is also provided to avail 2000 Hz as basis frequency.

5.

Spectrum 0Hz
-
150Hz

6.

Interference spectrum indicator


required on the front panel.

7.

Vector scanning


45 & 90 degrees,

8.

Mains supply


230 v ac./ 50Hz, Power consumption


50 watts

71.


MCPTR071

Tens unit
(transcutaneo
us electrical
nerve
stimulation)

1.

4 Channel
TENS with 8 electrodes

2.

Pulse width
-

50 microsec to 300 millisec

3.

Frequency
-

2 Hz to 600 Hz

4.

Intensity
-

0
-
60 mA

5.

Output voltage


0 to 50 Volts

72.


MCPTR072

Short wave
diathermy
(500 w)

1.

Continuous mode / pad electrode

2.

Input : 100
-
240 Volt AC,

3.

Frequency: 27.
12 MHz

4.

HF output: 200
-
500 W

5.

Treatment Time: 1
-
30 minutes with automatic shut off & buzzer

73.


MCMJ073

Automated
Blood Culture
Cystem

Fluorescent base alert system


74.


MCMJ074

Electronic Ph
Meter

Digital display


75.


MCMJ075

Research
Microscope

Binocular with (
5mp) Attachment to LCD Monitor


76.


MCMJ076

Diagnostic
Microscope
(Olympus)

Binocular Light Microscope


77.


MCMJ077

Electronic
Weighing
Machine

Able to weith o.5
-
20grams


78.


MCMJ078

Electronic
Centrifuge

Must have a speed regulator


79.


MCMJ079

Themo cycler

Autom
ated Programmable


80.


MCMJ080

Gel
Documentatio
n System

Standard System


81.


MCMJ081

Refrigerated
Centrifuge

For DNA


82.


MCDJ082

IOPA Dental
X
-
ray
Machine

Dental X
-
ray Should be capable of exposures tmj fronal maxillary sinus. It should be output of 55KV
to 85

KV, 10ma Generator should be 2.9V Expousure time 15 Secound or more Power requirement
Single Phase, 230 Volts, 50 Az, less than 15Amps.


83.


MCFJ083

Firearm
Weapons
Prototype

Pistol

Revolver

Airgun

Rifle


84.


MCFJ084

SLR digital
Camera with
access Ories

Expos
ure System

• Metering : Switchable: 40 segment honeycomb pattern; Center Weighted, Spot

• Exposure Compensation : ±3.0 EV (in 0.3EV, 0.5EV steps)

• Exposure Settings : Auto, Program Auto (P, with Program Shift), Shutter Priority (S), Aperture Prior

85.


MCB
J085

Refrigerated
Centrifuge (
-
20c) (20000
rpm)

• Max speed : 18,000 rpm.;

• Max rcf: 30,076 xg

• Max Volume: 1000 ml ( 4 x 250 ml)

• Temperature: 20 to Ambient, CFC free refrigeration

• Timer : Preselection of time up to 9 hrs. 59 min or continuous


Drive System : Maintenance Free, noiseless, bru


86.


MCBJ086

Spectrophoto
meter

UV visibledouble beam with multiple cuvette holders with attached PC, inbuilt stabilizer, two sets
of quartz cuvette, digital display with software control system .

Can perform k
inetic analysis


87.


MCBJ087

Spectro
Flourometer

PC controlled system with measurement modes for fluorescence, phosphorescence,
chemiluminescence and bioluminescence.

All reflective optical system with quartz over coated optics, Schwarzchild source optics fo
r increased
throughput.

Source: Xenon pulse la


88.


MCBJ088

Ultracenterifu
ge

Specification for Preparative Floor Model Ultracentrifuge equipped with a HEPA filter and necessary
rotors for Biosafety Lab

1. The ultracentrifuge should be capable of going to spe
eds of 1,00,000 RPM and forces up to
approximately 8,02,00 x g and should


89.


MCBJ089

Fully
Autoanayzer

SPECIFICATIONS FOR AUTOMATED BIOCHEMISTRY ANALYZER:

1. System: Fully automated, random access, floor model biochemistry analyzer with a minimum
throughp
ut of 800tests/hour , 400tests photometry/hr, 400ISE/hr

2. Reagents handling:

a. Accommodate a minimum


90.


MCBJ090

Automated
Electrophores
is

SPECIFICATIONS FOR FULLY AUTOMATED ELECTROPHORESIS SYSTEM:

1. The system should be fully automated electrophoresis
system based on Capillary electrophoresis
with 2 simultaneous migrations with complete walk away technology including migration and
quantita


91.


MCBJ091

Fully
Automated
Chemilumines
SPECIFICATION OF CHEMI
-
LUMINESCENCE

cence

Random Access Chemi
-
luminescence I
mmuno Assay Analyzer

1. System should be based on technology of paramagnetic particles with dio oxetebce phosphtas.

2. Atlast 22 assays from one sample.

3. Automated inventory managemen


92.


MCBJ092

Atomic
Absorption
Spectrop
hotomete

Fully controlled by PC,

Automatic wavelength Set
-
up, Automatic Slit switching, Automatic Lamp
current and gain set
-
up within 40 seconds. Auto switching and control 6 lamps on rotating turret,
Automatic flame ignition, automatic flow control for gases and

AUTOSAMPLER FOR FLAME
(OPTIONAL)

Up to 60 ea of 5ml vials, carousel type

Max. injection volume: 4ml

Accuracy :
-

2~3%

PRE REQUITES FOR AAS :
-


Description of goods/Qty

1. Two stage cylinder regulator brass chromplated body with S.S. Diap



93.


MCBJ093

Upright
Mic
roscope
(Complete
1.

General Specifications



System should consist of upright microscope with epi
-
fluorescent
with
accessories)

attachment, image acquisition unit (digital camera) and workstation
computer with data collection & processing software modules fo
r

• Bright field or Fluorescent

Karyotype Using Species Templates


For CGH, RxFISH and M
-
FISH cells the template must be attached before the cells are captured. For
bright field, fluorescent and standard FISH cells the template is attached as part of the
karyotyping
process. A classifie

FISH Software 8.

9. Other applications to be supported for upgradation 1. Multicolor/Multistain FISH (M
-
FISH)

2. Comparative Genomic Hybridization (CGH)

3. Color karyotyping
-
capability of karyotyping metaphase with FISH pro
bes; color caryotyping in
DAPI



94.


MCSKJ094

Phototherapy
Unit (U V A &
UVB COMBO)

a) Different combinations (UVA & NBUVA) can be provided.

b) Tubes from Philips Holland.

c) The Unit provided with imported mirror type reflectors.

d) Cabinet made of high qu
ality steel and powder coated.

e) Special UV chokes for maximum life for the tubes


95.


MCEJ095

Nasal & sinus
endoscopes
with CCD
camera
computer,
LED coldight
source, fibre
optic cable.
LED monitor
beam spliter

0,30,45,70 degree one ngid hopkins telescope 4
mm 18cm wide angle autocavable


96.


MCEJ096

Micromotor
drill with
handplace

MM drill with 40000 rpm. Angled hand piece conrolebox foot controle


97.


MCEJ097

Burrs set of 5
each
(imported) (a)
cutting
tonguston
carbide (b)
Diamond (c)
polishing

Cutting burrs
-

0.6mm to 6mm Set of 5
-

two set, diamondburrs
-

0.6 to 6mm set of five (two set)
polishing burrs (Size 6mm, 1mm, 2mm, 4mm, 6mm) tungustone carbide Length 70mm

98.


MCEJ098

Otoscope
-

Beta 200

Fibre opticullumination with seigle attachment.


99.


MCOPJ099

Operati
ng
Microscope

1. Unperellaled depth perception

2. X
-
Y coupling

3. Brilliant electronic.

4. fine focus adjustment and zoom through fooot switch

5. Effortless balancing

6. lllumination through fibre optic cable


100.


MCOPJ100

ND
-
yag laser
Nidek

1064mm Photodisr
uption


101.


MCOPJ101

Schiotz
Tonometer

Imported


102.


MCOPJ102

Fundus
Camera
(Zeiss)

3 in 1 multifunctional retinal camera lower flesh intencity

easu to use

continuous shooting function

auto focus and auto shoot


103.


MCORJ103

Mobile C
-
RAM (Image
Intenshifier
T.V
. System)

Technical Specifications

Generator Rating High Fequency

Frequency
-

40Khz

Power
-

3.5Kw 6.3 Kw for Rotating Anode

Radiography Mode

Kvp Output 40 to 100 Kvp 40 to 110 Kvp for Rotating Anode

mA Output
-

upto 50mA upto 80mA for Rotating Anode

mAs

Range



X
-

Ray tube
-


Focal Spot Stationary Anode
-

0.6/1.0mm 0.3/0.6mm for Rotating Anode

Optional
-

Line frequency models

Image Intensifier 9" Triple field, high gain, low dose image intensifier tube (optional 6" single / Due
field)

Camera
-

High sen
sitivity,



104.


MCORJ104

DHS & DCS
Instrument
Set with
Implants

Alu Case, blue, Yellow Lid, perforated, without contents

Upper Tray for Instrumenbts for 301.320, without Contents

Lower Tray for Instruments for 301.320, without Contents

DHS/ DCS Threaded Guid
e Wire, 2.5mm dia., L 230/5mm

DHS Angled Guide 135

DHS Angled


105.


MCORJ105

Digital
Pneumatic
Max pressure: 600 mmHg

Tourniquet

Voltage: AC 100
-
240V ~ 50/60Hz

Power consumption: 50W

Noise dB: below 60 dB

Time range: 5~180 min

Compressor weight: 5.6 kg

Compre
ssor dimension: 225(W) x 185(D) x 240 (H) mm

Built in lithium battery: life time 12~72 hours System fe


106.


MCORJ106

Surgical,
Diathermy
(bipolar)

§ The machine should be Microprocessor controlled with maximum output power of 400 W at 500
ohms

§ The modes of

operating should be monopolar and bipolar

§ Monoplar cut and coagulation. With coagulation having two modes : Contact and spray

§ Monopolar cu


107.


MCAJ107

Microscope

Student Type with Objectives
-

Scanning, Lowpower and Highpower, No Oil Immersion Lens, N
o
Built in Light Source


108.


MCAJ108

Microtone
Rotary with
Disposable
Blades

Rotary Microtone with Detachable Blades


109.


MCMDJ109

Video Upper
G.I.
Endoscope

specification gastrointestinal videoscope


Optical System Field of View : 145°

Direction of View

: 0° Forward Viewing

Depth of field : 3
-

100 mm

Distal End Outer : diameter

Specifications

Illumination : Lamp halogen or LED source

Switching of lamps : Manu
al.

Power supply : Voltage : PAL: 220
-
240V AC. NTSC: 100
-
120 V AC Frequency : 50/60 Hz

Size : Dimensions : 295(W) x


110.


MCMDJ110

Video
Colomoscope

Specifications

Illumination : Lamp halogen or LED source


Switching of lamps : Manual.

Power supply : Voltage : PAL: 220
-
240V AC. NTSC: 100
-
120 V AC Frequency : 50/60 Hz

Size : Dimensions : 295(W) x

Specifications

Illumination : Lamp halogen or LED source


Switching of lamps : Manual.

Power supply : Voltage : PAL: 220
-
240V AC. NTSC: 100